Welcome: chuangen medical
Language: Chinese ∷  English
chuangen.medical@aliyun.com +86 0556-5222292

CN-GC112A Gas Chromatograp

Features

* Brand-new designed microcomputer temperature control system of world advanced control technology. High accuracy of temperature control (optimum: ± 0.05°C), high reliability and anti-interference.



* Control and operating parameters (including for detectors) set up by the keyboard. Self-diagnosis system, power-fail protection, file storage and call, limited temperature setting. Accurate display of setting and actual value of temperature control of each path, bridge current of TCD, sensitivity of FID, retention and analysis time.



* Four paths of independent control sections: column oven, sample injector, detector and thermal conductive pond.



* 5-ramp programmed temperature rising control. Auto-drop of temperature of column oven (auto-opening the door behind the oven) to a fast cooling.



* Dual FID provided with options of GC112 A-TCD, GC112A-ECD, GC 112 A-FPD and GC112A-NPD, gas sampling valve, transformation oven, absorber and pyrolysis absorber available for selection. Two detectors can be installed at the same time with high sensitivity. Short time to be stabilized and easy to clean and install nozzle.



* Packed column and capillary column (0.53 large diameter or small diameter) can be installed easily including a complete set of split sampling connector and tailing connector to be installed free of charge. Optional 0.53mm large diameter capillary direct sampler is available for choice.



* Analysis can be conducted in packed or capillary column with multiple sampling methods. Packed column: on-column sampling, instantaneous vaporized sampling and gas sampling. Capillary column: split sampling and 0.53mm large diameter capillary direct sampling.



* Construction of dual gas path, dual sample injector and dual packed column system can effectively avoid baseline undulation and drift in real time and facilitate programmed temperature rising analysis. Accurate and scaled type gas path control valves will provide a high repeatability and stability.



* Advanced functions: auto ON/OFF temperature control system, permanent data storage, file storage and call and over-temperature protection.



* Large volume column (300x280x270) can hold capillary column and dual packed column. Built-in heating filament structure is in favor of quartz capillary analysis.



* A complete set of accessories including purifier, pressure-reducing valve, gas path tube, wrench and tools, injection needle and connectors of all kinds are provided free of charge and are ready to connect with supply-air system prepared by users.

Features

* Brand-new designed microcomputer temperature control system of world advanced control technology. High accuracy of temperature control (optimum: ± 0.05°C), high reliability and anti-interference.

 

* Control and operating parameters (including for detectors) set up by the keyboard. Self-diagnosis system, power-fail protection, file storage and call, limited temperature setting. Accurate display of setting and actual value of temperature control of each path, bridge current of TCD, sensitivity of FID, retention and analysis time.

 

* Four paths of independent control sections: column oven, sample injector, detector and thermal conductive pond.

 

* 5-ramp programmed temperature rising control. Auto-drop of temperature of column oven (auto-opening the door behind the oven) to a fast cooling.

 

* Dual FID provided with options of GC112 A-TCD, GC112A-ECD, GC 112 A-FPD and GC112A-NPD, gas sampling valve, transformation oven, absorber and pyrolysis absorber available for selection. Two detectors can be installed at the same time with high sensitivity. Short time to be stabilized and easy to clean and install nozzle.

 

* Packed column and capillary column (0.53 large diameter or small diameter) can be installed easily including a complete set of split sampling connector and tailing connector to be installed free of charge. Optional 0.53mm large diameter capillary direct sampler is available for choice.

 

* Analysis can be conducted in packed or capillary column with multiple sampling methods. Packed column: on-column sampling, instantaneous vaporized sampling and gas sampling. Capillary column: split sampling and 0.53mm large diameter capillary direct sampling.

 

* Construction of dual gas path, dual sample injector and dual packed column system can effectively avoid baseline undulation and drift in real time and facilitate programmed temperature rising analysis. Accurate and scaled type gas path control valves will provide a high repeatability and stability.

 

* Advanced functions: auto ON/OFF temperature control system, permanent data storage, file storage and call and over-temperature protection.

 

* Large volume column (300x280x270) can hold capillary column and dual packed column. Built-in heating filament structure is in favor of quartz capillary analysis.

 

* A complete set of accessories including purifier, pressure-reducing valve, gas path tube, wrench and tools, injection needle and connectors of all kinds are provided free of charge and are ready to connect with supply-air system prepared by users.

 

* Temperature control

Control range

7°C-400°C above room temperature  (increment of 1°C)  

Control objects

column oven, detector, thermal conductive pond and sample injector

Programmed temperature rising

5 ramps

Rate of programmed temperature rising

0.1°C -40°C (increment of 0.1°C)

Thermostatic time

0-655 min (increment of 1 min)

 

* FID

Sensitivity (minimum testability)

M£5´10-11 g/s (sample: C16)

Optimum testing result

M£1´10-11 g/s

Noise

£5´10-14 A

Drift

£6´10-13 A/h

Linear range

³106

 

 

* TCD 

Sensitivity

³2000mV.ml/mg

Noise

£20mV

Drift

£30mV

Linear range

³104

 

* ECD 

Temperature range

350°C Max

Measuring mode

63Ni

Sensitivity

£10-13 g (Benzex)

Linear range

³103

 

* FPD 

Sensitivity (minimum testability)

£2´10-11 g (P)/s (sample P in the parathion methyl)

Optimum testing result: £3.8´10-12 g (P)/s.

Sensitivity (minimum testability)

£1´10-10 g (S)/s (sample: S in the parathion methyl)

Optimum testing result: £2.7´10-11 g (S)/s

Linear range

³103 (sample P parathion)

³102 (sample S parathion)

Stability

£30mV/h

Noise

£10mV

Starting time

not greater than 2 hours

 

* NPD

Sensitivity (minimum testability)

Dt(N)£5pg (N)/s (sample: azobenzene)

Optimum testing result: 0.5ppm

Sensitivity (minimum testability)

Dt(P) £0.5pg (P)/s (sample: malethion)

Optimum testing result: 0.2ppm

Drift

£8´10-13 A/h

Noise

£5´10-13 A

Linear range

³103

Max application temperature

360°C


INQUIRY

CATEGORIES

CONTACT US

Contact: Katrina Yao

Phone: +0086 18255652021

Tel: +86 0556-5222292

Email: chuangen.medical@aliyun.com

Add: Area C, Huamao 1958 Project, No.80, South Fangzhi Road, Daguan District, Anqing, Anhui, China.